Skip to content

purvigor KN95 Mask particle filter mask — health risk

EU Safety GateSerious riskOther health riskDec 25, 2020
Photo from the official notice: purvigor KN95 Mask particle filter mask — health risk
Photo from the official notice: purvigor KN95 Mask particle filter mask — health risk
Photo from the official notice: purvigor KN95 Mask particle filter mask — health risk
Photo from the official notice: purvigor KN95 Mask particle filter mask — health risk
Photo from the official notice: purvigor KN95 Mask particle filter mask — health risk

Photos from the official notice: EU Safety Gate (European Commission) · image rights

Hazard

The particle/filter retention of the material is insufficient (measured value 64,6%). Consequently, an excessive amount of particles or microorganisms might pass through the mask, increasing the risk of infection if not combined with additional protective measures. The product does not comply with the Personal Protective Equipment Regulation and with the relevant European standard EN 149.

Measures taken

  • Recall of the product from end users — Importer (14/12/2020)

Product description

Single use particle half-face filter mask for protection against particles and aerosols, labelled KN95 and FFP2. The brand's name is embossed on the side.

What should I do?

If you own this product, stop using it and check the official notice for what the seller or authority advises (return, refund, repair or disposal). Measures listed here were reported by national authorities to the EU Safety Gate system.

Recent alerts in this category

See all →