TRANS MEDIKAL - GETINGE steam generator — burns risk
Photos from the official notice: EU Safety Gate (European Commission) · image rights
Hazard
Due to a manufacturing error, excess pressure can build up in the steam generator and a significant amount of energy could be released resulting in injuries, including burns. The product does not comply with the requirements of the Pressure Equipment Directive.
Measures taken
- Recall of the product from end users — Distributor (02/03/2025)
- Making the marketing of the product subject to prior conditions — Distributor (02/03/2025)
Product description
Steam generator integrated into steriliser.
What should I do?
If you own this product, stop using it and check the official notice for what the seller or authority advises (return, refund, repair or disposal). Measures listed here were reported by national authorities to the EU Safety Gate system.