JMF Dermapen skin treatment device — electric shock risk

EU Safety GateSerious riskElectric shockMar 13, 2015
Photo from the official notice: JMF Dermapen skin treatment device — electric shock risk
Photo from the official notice: JMF Dermapen skin treatment device — electric shock risk

Photos from the official notice: EU Safety Gate (European Commission) · image rights

Hazard

The power supply does not have adequate insulation and is not designed to withstand foreseeable overloads. The user may consequently receive a mains voltage electric shock. The lack of shutters on the plug adaptor may result in the user receiving an electric shock when plugging / unplugging the appliance. The product does not comply with the requirements of the Low Voltage Directive and the relevant European standard EN 61558 and the relevant national standard.

Measures taken

  • Destruction of the product — Other (26/02/2015)

Product description

Hand held pen style appliance with many tiny needles for skin treatment, supplied with a mains transformer and a control box, accompanied by a plug in adaptor to convert the fitted 2 pin plug to a UK 3 pin configuration. Packaged in plain white cardboard box.

What should I do?

If you own this product, stop using it and check the official notice for what the seller or authority advises (return, refund, repair or disposal). Measures listed here were reported by national authorities to the EU Safety Gate system.