BEHA-AMPROBE Multimeters — fire risk

UK OPSSMedium riskFireMar 17, 2023
Photo from the official notice: BEHA-AMPROBE Multimeters — fire risk

Photos from the official notice: UK Office for Product Safety and Standards (GOV.UK) · image rights

Hazard

The product has been recalled as it was discovered that an improperly rated fuse for CAT III 600V applications was installed in the affected products. The breaking capacity of the installed fuse is insufficient for the prospective fault current under foreseeable misuse conditions (abnormal application of electrical installation voltage above 250 volts between A and COM terminals) and may lead to venting or rupture of the fuse and the subsequent possibility of arc explosion. An arc explosion within the product enclosure may readily escape and result in a fire or severe bodily harm. The product does not meet the requirements of the Electrical Equipment (Safety) Regulations 2016

Remedy

Recall of the product from end users. For more information, please access the BEHA-AMPROBE recall site at the following link: https://www.beha-amprobe.com/en/announcements/

Product recall

Measures taken

  • Recall of product from end users

Product description

The products are red/black/grey digital multi meters used for making electrical measurements

What should I do?

If you own this product, stop using it and follow the corrective action above: contact the retailer or supplier for a repair, replacement or refund. This notice was published by the UK Office for Product Safety and Standards.