TRUMPF Medical Systems Recalls Television Mounting System Due to Falling Hazard

Photos from the official notice: U.S. Consumer Product Safety Commission · image rights
Hazard
The arm joints on the television mounting system do not have a washer at the top joint which can result in excessive wear of the stop pins on the arm. This can allow the arm system and items connected to the arm to fall and injure the user or bystanders.
Remedy
Secure the unit to prevent movement and contact TRUMPF Medical Systems for information on receiving a repair kit that includes the missing washers.
Repair
Incidents & injuries
The firm received one report of a mounting system falling while being positioned. No injuries have been reported.
Product description
The mounting systems were sold for use in hospital Intensive Care Units. The units are off-white and have multiple joints between the ceiling mount and the TV mount at the end of the arm system.
Sold on
TRUMPF Medical Systems Inc. nationwide From January 2005 through December 2011 for about $2,100.
Contact the company
For additional information, contact TRUMPF Medical Systems Inc. toll-free at (888) 474-9359 between 8 a.m. and 5 p.m. ET Monday through Friday, or email [email protected]
What should I do?
Stop using the product and follow the remedy below. Contact the company for a refund, repair or replacement — recall remedies are free.
Recent alerts in this category
- Cambridge Audio Ceiling Speaker Fire Hood — fire riskUK OPSSSep 16, 2026· Cambridge Audio· Communication and media equipment
- FocusRay Wired Web Cam — fire riskUK OPSSSep 15, 2026· FocusRay· Communication and media equipment
- ALEENFOON Caregiver Pager — fire riskEU Safety GateSeriousSep 11, 2026· Aleenfoon· Communication and media equipment
- KOSS On-ear headphones — environment riskEU Safety GateSeriousSep 4, 2026· KOSS· Communication and media equipment
- Audio Partnership Recalls Cambridge Audio Yoyo (M) Portable Bluetooth Speakers Due to Fire HazardU.S. CPSCSep 3, 2026· Audio Partnership· Communication and media equipment
- Mini Projector Q5 Mate — electric shock risk (SR/02336/26)EU Safety GateSeriousAug 28, 2026· Communication and media equipment