Skip to content

Extech Instruments Recalls Digital Clamp and Multimeters due to Electrocution Hazard

U.S. CPSCElectric shockDec 14, 2010
Clamp Meter
Clamp Meter
Multimeters
Multimeters

Photos from the official notice: U.S. Consumer Product Safety Commission · image rights

Hazard

When the battery runs low, the meters can fail to give an accurate voltage reading, resulting in the operator falsely believing the electrical power is off or low. This poses an electrocution hazard.

Remedy

Consumers should immediately stop using the recalled digital clamp meters and multimeters and contact Extech for a free replacement meter.

Replace

Incidents & injuries

Extech received one report of a meter displaying an incorrect voltage reading. No injuries have been reported.

Product description

This recall involves Extech digital clamp meters with model numbers EX612, EX613, EX622 and EX623 and multimeters with model numbers EX540, EX542 and EX570. "Extech" and the model number can be found on the front of the product and the serial number on the rear. Serial numbers that begin with "A" are not included in the recall.

Sold on

Industrial and electrical distributors and wholesalers nationwide and at online tool and test equipment retailers from January 2008 to November 2010 for between $150 and $300.

Contact the company

For additional information, contact Extech toll-free at (855) 239-8324 between 8:30 a.m. and 5 p.m. ET Monday through Friday or visit the company's website at http://www.extech.com/recall

What should I do?

Stop using the product and follow the remedy below. Contact the company for a refund, repair or replacement — recall remedies are free.