700HPM High Pressure Calibration Manifold — injuries risk


Photos from the official notice: UK Office for Product Safety and Standards (GOV.UK) · image rights
Hazard
This product presents a serious risk of injuries as during manual pressurization of the equipment, the plunger nut threads of the High-Pressure Pneumatic Calibration Manifold may yield releasing the plunger assembly and allowing it to be ejected at high velocity. The threads securing the plunger assembly have been found to fail prematurely. This can result in serious injury to the user. The product does not meet the requirements of the Pressure Equipment (Safety) Regulations 2016.
Remedy
Fluke UK Ltd has stopped shipping the affected manifold and recalled the product from end users. Further information is available on the Fluke website at the following link: https://www.fluke.com/en-gb/support/safetynotices/fluke-700hxx-recall .
Product recall
Measures taken
- Recall of product from end users
Product description
700HPM High Pressure Calibration Manifold.
What should I do?
If you own this product, stop using it and follow the corrective action above: contact the retailer or supplier for a repair, replacement or refund. This notice was published by the UK Office for Product Safety and Standards.
More from this brand
- Fluke Test Lead - Model 3PHVL-17XX — electric shock riskUK OPSSJul 7, 2023· Fluke· Electrical appliances and equipment
- Fluke 8x V Series Digital Multimeters (DMM) — electric shock riskUK OPSSFeb 16, 2023· Fluke· Electrical appliances and equipment
- Noncontact Electrical Tester Recalled by Fluke Due to Shock or Burn HazardU.S. CPSCMar 11, 2010· Fluke· Electrical appliances and equipment
- Digital Clamp Meters Recalled by Fluke Due to Shock HazardU.S. CPSCMay 15, 2009· Fluke· Electrical appliances and equipment
- CPSC, Fluke Corp. Announce Recall of Electrical Testing ComponentsU.S. CPSCMay 3, 2004· Fluke· Electrical appliances and equipment
- CPSC, Fluke Corp. Announce Recall of Digital MultimetersU.S. CPSCAug 19, 2002· Fluke· Electrical appliances and equipment
Recent alerts in this category
- PCL Push Button Safety (PBS) Coupling — injuries riskUK OPSSJul 24, 2026· PCL· Pressure equipment/vessels
- Ultra Space Tank nitrous oxide container — injuries riskEU Safety GateSeriousMay 8, 2026· Ultra Space· Pressure equipment/vessels
- Rosenbauer water cannon/monitor — injuries riskEU Safety GateSeriousOct 31, 2025· Rosenbauer· Pressure equipment/vessels
- Jungeblodt pressure valve — injuries riskEU Safety GateSeriousOct 10, 2025· Jungeblodt· Pressure equipment/vessels
- Sempell pressure valve — injuries riskEU Safety GateSeriousOct 10, 2025· Sempell· Pressure equipment/vessels
- Cochran Ltd Ultranox Boiler Burner — fire riskUK OPSSAug 4, 2025· Ultranox· Pressure equipment/vessels